Print

Microbiological safety cabinet – series NC

Personnel, product and environment protection
Protective class II - SIST EN 12469:2000 and DIN 12980:2005
Class II, Type B2 - NSF: 2008;
Working area GMP class A - class 5 by 14644-1:1999

NC

CEZaščita proizvodaZaščita okoljaZaščita operaterja

The bench has 100% in/out air current. The exhaust  air has to direct connected to the independent AHU system with additional HEPA filter. Two fans inside bench  and filtrs system assure safety conditions for product and environment.

Basic characteristics:

  • easy maintenance;
  • easy cleaning and disinfecting;
  • rounded corners;
  • easy and safe filters replacement;
  • integrated low-noise fan;
  • automatic height placement of the front safety glass barrier;
  • available in different sizes;
  • easily accessible electrical components;
  • 100% in/out air

Application fields:

  • microbiology;
  • pharmacy;
  • biotechnology;
  • medicine;
  • clinics and hospitals;
  • preparation of cytotoxic agents, antibiotics;
  • laboratories for research and development.

TypeMC 9-NCMC 12-NCMC 15-NCMC 18-NC
External dimensions
  w x d x h [mm]
1030x990x2220 1340x990x2220 1650x990x2220 1920x990x2220
Working area
w x d x h [mm]
880x600x600 1190x600x600  1500x600x600  1810x600x600 
Maximal power consumption (W) 2500 2500  2500  2500 
Maximal power consumption on socket (W)  550  550  550 550 
Laminar air flow (m/s) 0,4  0,4  0,4 0,4 
Voltage (V/Hz) 230/50  230/50 230/50  230/50 
Weight [kg] 260  305 365 440
ABS. working filter H14 305x610x69mm [piece] 3 4 5 6
ABS. outlet filter H14 305x457x69mm [piece] / 2 / 3
ABS. outlet filter H14 305x457x78mm [piece] / / 2 /
ABS. outlet filter H14 457x457x69mm [piece] 1 / / /
ABS. outlet filter H14 [piece] 3 5 6 7