Print

Measuring the speed of air flow

We perform measurements of speed of the air flow in the work area of chamber and the inlet air flow speed. The test is performed in accordance with the recommendations of standard EN 12469:2000 and cGMP Annex 1.

 Measuring the speed of air flow

Before the test we arrange measuring points logically, the method of measurement depends on the type of protective chamber. Speed must be within the acceptance criteria. In the case of excessive deviation, we set operating parameters of the chamber, if possible. The test is performed in accordance with the recommendations of standard EN 12469:2000 and cGMP Annex 1.

We use measuring equipment from renowned manufacturers Testo, Schiltknecht, TSI.

All instruments are calibrated annually.